Chapter 9: Maintenance
Pulse Oximeter for Sleep Screening Operation Manual 9-1
Chapter 9: Maintenance
Schedule of Maintenance
MAINTENANCE: HOW OFTEN: BY DOING THIS:
Replace battery When Low Battery Indicator
is flashing or 12-15 hours of
battery use has elapsed.
Follow the instructions for
installing the batteries.
Disinfect the reusable sensor Before attaching the sensor
to the patient.
Follow the instructions for
cleaning the reusable sensor.
Disinfect the monitor When necessary. 1. Remove the batteries from
the unit.
NOTE: print tend data
before cleaning. Trend
data will be lost if batteries
are disconnected more
than one and a half
minutes.
2. Wipe the surfaces of the
monitor with a soft, clean
cloth dampened in
isopropyl alcohol. Use only
a cloth that is dampened,
not wet.
2 CAUTION! Do not
allow isopropyl alcohol or
any liquid to enter any of
the openings on the
monitor
.
Storage
Whenever possible, the monitor should be stored at room temperature in a dry environment.
If it is necessary to store the monitor for an extended period of time, the unit should be packed
in its original shipping container. Storing the monitor for a long period of time may degrade
the battery capacity. Batteries should be removed from the monitor before storing.
Storage specifications are as follows:
Temperature: -40
o
C to 75
o
C (-40
o
F to 167
o
F)
Relative Humidity: 10% to 95% (noncondensing)