Warnings, Cautions, & Notes
vi Pulse Oximeter for Sleep Screening Operation Manual
NOTE! Dyes introduced into the bloodstream, such as methylene blue, indocyanine
green, indigo carmine, patent blue V (PBV), and fluorescein, may adversely affect the
accuracy of the SpO
2
reading.
NOTE! Any condition that restricts blood flow, such as use of a blood pressure cuff or
extremes in systemic vascular resistance, may cause an inability to determine accurate
pulse rate and SpO
2
readings.
NOTE! Remove fingernail polish or false fingernails before applying SpO
2
sensors.
Fingernail polish or false fingernails may cause inaccurate SpO
2
readings.
NOTE! Batteries are designed to be replaceable by individuals trained in this procedure.
Follow local governing ordinances and recycling instructions regarding disposal or
recycling of device components, including batteries.
NOTE! SpO
2
averaging is the number of pulse beats over which the SpO
2
value is
averaged; pulse averaging is the number of seconds over which the pulse value is
averaged.
NOTE! Hazards arising from software errors have been minimized. Hazard analysis was
performed to meet EN60601-1-4:2000.
NOTE! Optical cross-talk can occur when two or more sensors are placed in close
proximity. It can be eliminated by covering each site with an opaque material.
NOTE! All user and patient accessible materials are non-toxic.